The Solitax highline sc turbidity and suspended solids insertion probe provides high-precision results in clear to strongly turbid liquids as well as in high concentrated sludges. Since this probe is for in pipe use only an appropriate mounting assembly for insertion application is required. Turbidity measurement performance is certified to MCERTS regulation and the probe operates with a 90° measuring method according to DIN EN ISO 7027. The suspended solids measurement by a colour-independent method with an infrared duo scattered light photometer is in accordance with DIN 38414. A proven automatic wiper cleaning system allows sustained performance quality in drinking water as well as in waste water applications. The Solitax highline sc probe material is stainless steel and measurement data is displayed and processed with the help of a digital Hach SC200 or SC1000 controller.
Furthermore, the Solitax highline sc turbidity immersion probe is compatible for Prognosys which is a predictive diagnostic system that allows you to be proactive in your maintenance, by alerting you to upcoming instrument issues. Know with confidence whether changes in your measurements are due to changes in your instrument or your water.
Please contact Hach for further information. Request for a quote today!
This instrument connects to Claros, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations. To unlock the full potential of Claros, insist on Claros Enabled instruments.
| Accuracy | Turbidity up to 1000 NTU: without calibration < 5% of the measured value ±0.01 NTU with calibration < 1% of the measured value ±0.01 NTU |
| Automatic Wiper | Yes |
| Body Material | Stainless Steel |
| Cable Length | 10 m (33 ft) (optional extension cables available) |
| Calibration Method | Turbidity: Formazin or Stablcal Standard (at 800 NTU). Suspended Solids: Sample specific, based on gravimetric TSS analysis with a correction factor procedure. |
| Certifications | CE, MCERTS |
| Controller | Sensor only |
| Controller Compatibility | SC4500, SC1000, SC200 |
| Diameter Sensor | 60 mm |
| Flow | Max. 3 m/s (the presence of air bubbles affects the measurement) |
| IP Rating | IP68 |
| Length | 315 mm |
| Maintenance Interval | 1 h/month |
| Material | Optics carrier and sleeve: stainless steel 1.4571 Wiper arm: stainless steel 1.4581 Wiper rubber: silicone rubber (standard) Optional: FKM/FPM (LZX578) Wiper shaft: stainless steel 1.4571 |
| Max. cable length Sensor-to-Controller | 100 |
| Measuring Principle | Infrared Duo scattered light technique for color-independent turbidity measurement Turbidity in accordance with DIN EN ISO 7027 / TS equivalent DIN 38414 |
| Model | Solitax highline sc |
| Mounting | Insertion |
| Operating Temperature Range | 0 – 40 °C (32 – 104 °F) |
| Parameter | High Range Suspended Solids, Turbidity |
| Pressure Range | Stainless steel insertion sensor: <6 bar or <60 m |
| Range | TSS content: 0.001 – 500 g/L; 0.001 – 500,000 mg/L (the upper range is dependent on the sample property) Turbidity: 0.001 – 4000 NTU |
| Repeatability | TSS content: < 3%|Turbidity: < 1% |
| Response Time T90 | 1 – 300 s adjustable |
| Special Notes | Installation note: Distance sensor-wall > 50 cm (Turbidity) |
| Warranty | 12 months |
| Weight | Insertion stainless steel: 2.4 kg (5.29 lb.) |
| What’s included? | Turbidity/Suspended Solids probe, user manual |

Tiếng Việt