Smart amperometric Ozone Analyser: 9185 sc
The 9185sc Ozone Sensor is an online, single-channel industrial sensor that measures ozone in drinking water treatment plants, distribution networks, and other applications that require monitoring of ozone at the ppb and ppm levels. This instrument uses an amperometric method to measure ozone concentration. A membrane allows the selective diffusion of ozone molecules to the amperometric sensor. The measurement is compensated for pH and temperature.
| Accuracy | 3 % or ± 10 ppb O3, whichever is greater |
| Calibration Interval | 2 months |
| Calibration Method | Zero Calibration: Electrically or with ozone-free water Calibration: Comparison of lab method with process sample |
| Enclosure Rating | IP66 / NEMA 4X |
| Flow | 14 L/h (200 – 250 mL/min) auto-regulated by flow through cell |
| Interferences | No interference from bromine, chloramines, chlorine, chlorine dioxide, or hydrogen peroxide |
| Lower Limit of Detection (LOD) | 0.005 mg/L ozone (O3) |
| Maintenance Interval | 6 months for membrane and electrolyte (measurement cell) |
| Measuring Interval | Continuous |
| Measuring Principle | Amperometric/Membrane (electrode, membrane, electrolyte) |
| Mounting | Flat, vertical surface (panel, stand, etc.) |
| Operating Temperature Range | 0 – 45 °C |
| Parameter | Ozone |
| Pressure Range | 0.1 – 2 bar in flow cell |
| Range | 0.005 – 2 mg/L O3 |
| Relative Humidity | 0 – 90 % non-condensing |
| Response Time | 90 % in less than 90 seconds |
| Sample Input | 0.25 ” outer diameter |
| Sample Output | 0.5 ” inner diameter |
| Sample Temperature | 2 – 45 °C |
| Standard deviation | 1.0 |
| Storage Conditions | -20 – 60 °C |
| Temperature Compensation | Automatic over sample temperature range |
| Temperature Compensation | Automatic over sample temperature range |
| Warranty | 12 months |
| Weight | 6.5 kg |
| What’s included? | Hach 9185sc Ozone sensor panel, 4 spare membranes, and Manual. |