The first maintenance-free optical oxygen sensor for power plants.
High accuracy Luminescent Dissolved Oxygen (LDO) Sensor optimised for power industry application.
The absence of membrane and electrolyte means that the sensor accuracy is unaffected by process changes such as changes in flow. Maintenance and operating costs are significantly reduced. The sensor has been designed to ensure mechanical robustness to extend operational lifetime and optimise its total cost of ownership.
The quick response time comes from the Orbisphere K1100 two second measurement frequency. Capable of measuring accurately at this frequency over a 12 month period. With no calibration required, the K1100 surpasses other optical and electrochemical sensors that display significant drift after only a few months in the same conditions. This optical sensor is designed for minimal drift, resulting in it being the most stable sensor with the longest calibration intervals achievable in the market. This is possible by its long-life spot and optimised controller software.
Maintenance intervention is limited to 2 minutes and a zero point calibration, offering significant cost benefits compared with traditional electrochemical sensors and other luminescent sensors. Using gas phase calibration means chemicals are not required, and therefore the task is easier and safer without reducing measurement precision.
| Accuracy | ± 0.02 ppm or 3 % whichever is greater |
| Ambient Temperature | -5 – 50 °C (23 – 122°F) |
| Analog Outputs | 3 Smart 0/4 to 20 mA (500 Ohms) programmable as linear or tri-linear, configurable to send diagnostics or alarm informations. |
| Application | In-line |
| Calibration | Two points at cap replacement (zero and air), one during use (air) |
| Certifications | 2004/108/EC – EN 61326-1 |
| Communication Capabilities | 3 x 0/4-20 mA; RS485; Ethernet |
| Display Resolution | 0.1 ppb |
| Lower Limit of Detection (LOD) | 0.015 ppm |
| Mounting | 28 mm Orbisphere fitting |
| Parameter | Oxygen |
| Power Requirements (Voltage) | 100 – 230 VAC |
| Range | 0 – 40 ppm dissolved O2 (DO) |
| Relays | Measurement board: 3 measurement alarm relays (1A-30 VAC or 0.5A-50 VDC), configurable to send diagnostics information. Main board: 1 system alarm relay (1A-30 VAC or 0.5A-50 VDC). |
| Repeatability | ± 0.015 ppm or 2 % whichever is greater |
| Reproducibility | ± 0.02 ppm or 3 % whichever is greater |
| Response Time | (90%) <10 s (gas phase); <50 s (liquid phase) |
| Sample Flow Rate | 50 – 300 mL/min |
| Sample Pressure | 1 – 20 bar absolute (14.5 – 290 psi) |
| Sample Temperature | -5 – 50°C (23 – 122°F) |
| Sensors | K1100 LDO |
| Storage Conditions | -5 °C – 100 °C |
| Temperature Range | Accurate from -5 – 50 °C (23 – 122 °F) Resistant – temperature from -5 – 100 °C (23 – 212 °F) |
| Warranty | 12 months |
| Weight Sensor | 0.6 |
| What’s included? | K1100 Sensor only Controller, sensor cable, flow-chamber or process connection must be ordered separately |