Using the Nitratax eliminates reagents, sampling, and sample conditioning. The UV absorption method is proven, continuous, and precise. The probe is self-cleaning. Life-long factory calibration and full-featured “Plug & Play” with SC Controllers; Nitratax must be connected to an SC controller to form a complete system.
Prognosys is a predictive diagnostic system that allows you to be proactive in your maintenance, by alerting you to upcoming instrument issues. Know with confidence whether changes in your measurements are due to changes in your instrument or your water.
This instrument connects to Claros, Hach’s innovative Water Intelligence System, enabling you to seamlessly connect and manage instruments, data, and process – anywhere, anytime. The result is greater confidence in your data and improved efficiency in your operations. To unlock the full potential of Claros, insist on Claros Enabled instruments.
| Accuracy | ± 5 % of measured value + 0.5 mg/L (with standard solutions) |
| Ambient Temperature | 2 – 40 °C (36 – 104 °F) |
| Application | Drinking Water |
| Cable Length | 10 m (32.8 ft.), fixed cable at sensor |
| Controller | Sensor only |
| Controller Compatibility | SC200, SC1000 |
| Diameter | 75 mm |
| Includes | Nitrate probe, standard solution, user manual |
| Length | 327 mm |
| Lower Limit of Detection (LOD) | 0.5 mg/L to 20 NO₂₊₃–N (with standard solutions) |
| Maintenance Requirements | 1 hour/month typical |
| Measuring Interval | 5 to 30 min (fixed values selectable) |
| Measuring Principle | UV absorption measurement |
| Model | Nitratax clear sc |
| Operating Pressure | 0.5 bar (7.2 psi), maximum |
| Operating Temperature Range | 2 – 40 °C |
| Path Length | 5 mm |
| Power Consumption | 2 |
| Pressure Range | ≤ 0.5 bar |
| Range | 0.5 – 20 mg/L NO2+3-N 2 – 100 mg/L NO3 |
| Range | 0.5 – 20 mg/L NO₂₊₃-N |
| Resolution | 0.1 mg/L |
| Response Time | 5 min |
| Sample Flow Rate | 0.5 – 10 L/h sample |
| Service Interval | 6 months or as experience dictates |
| Sludge Compensation | No |
| Verification | Using standard solutions |
| Warranty | 12 months |
| Weight | 3.3 kg |
| What’s included? | Nitrate probe, standard solution, user manual |