The Phosphax sc Analyzer is ideal for precise phosphate measurement in rugged outdoor environments. Perfect for a wide range of wastewater and drinking water applications, it features a broad measurement range, minimal maintenance, and predictive diagnostics for optimal performance.
The Phosphax sc Phosphate Analyzer is designed for phosphate measurement and phosphate analysis in environmentally controlled, rugged outdoor installations, with the ability to measure phosphate across a wide range of wastewater applications, from 0.05 to 15.0 mg/L. It offers a broad measurement range suitable for various wastewater and drinking water applications, featuring a self-cleaning membrane that requires minimal maintenance.
Phosphax sc must be used with a Hach controller and is compatible with Hach SC4500 or SC1000.
Prognosys, a predictive diagnostic system, allows you to proactively manage maintenance by alerting you to upcoming instrument issues. This ensures you can confidently determine whether changes in your measurements are due to the instrument or the water.
| Accuracy | Using standard solutions: 2% ± 0.05 mg/L |
| Cable Length | 2 m fixed power cable at analyzer |
| Cable Length | 2 m fixed data cable at analyzer |
| Channels | 1 |
| Data Transfer | Current output, relays and bus communication via SC controller |
| Dimensions (H x W x D) | 720 mm x 540 mm x 390 mm |
| Enclosure Rating | IP55 |
| Filter Configuration | Use e.g. Filtrax |
| Flow | 1 – 20 L/h Sample (free of suspended solids) |
| Fuse Protection | Via SC1000 controller or power box. Maximum of 2 analysers per SC1000 or power box |
| Housing Type | Outdoor |
| Indoor/Outdoor | Outdoor |
| Lower Limit of Detection (LOD) | 0.05 mg/L PO4-P |
| Material | ASA/PC UV-resistant |
| Measurement Method | Photometric method using vanadate-molybdate. |
| Measuring Interval | 5 – 120 min, adjustable per 5 min. |
| Measuring Principle | Phosphate ions react with vanadate-molybdate reagent to form a yellow dye. |
| Number of Channels | 1 |
| Operating Temperature Range | -20 – 45 °C |
| Outputs | Current output, relays and bus communication via SC controller |
| Parameter | ortho-Phosphate |
| Permissible Chloride Range | Max. Cl– concentration: 1000 mg/L |
| pH Range | 5 – 9 pH |
| Power Requirements (Hz) | 50 Hz |
| Power Requirements (Voltage) | 115 – 230 VAC, powered by Power Box or SC Controller |
| Power Supply | 115-230 VAC |
| Pressure Range | -30 – 50 mbar with continuous sample preparation; at overflow vessel |
| Range | 0.05 – 15.0 mg/L PO₄-P |
| Reagent Consumption | 500 mL/month |
| Reproducibility | 2 % + 0.05 mg/L |
| Response Time | < 5 min |
| Sample Delivery Hose Options | Outdoor |
| Sample Level | Level of liquid in basin with filtration probe must be below analyzer |
| Sample Quality | Ultra filtrated or comparable |
| Sample Temperature | 4 – 40 °C (39 – 104 °F) |
| Sampling Option | 1 channel continuous sample from filtration unit, e.g. Filtrax |
| Storage Conditions | -20 – 60 °C (-4 – 140 °F), 95% relative humidity, non-condensing |
| Warranty | 12 months |
| Weight | Without sample preparation system and without chemicals: 31 kg |
| What’s included? | 0.05 – 15 mg/L PO₄-P, 115-230 VAC Phosphax sc Ortho-Phosphate Analyzer, for one channel continuous sample, 3 month supply of all necessary consumables at 5 minute cycle time, user manuals |