Smart amperometric Total Free Chlorine Sensor: 9184sc
The low minimum detection limit and wide measurement range of the Hach 9184sc Amperometric Free Chlorine Sensor makes it ideal for efficient residual chlorine monitoring. Continuously measures hypochlorous acid (HOCl), Hypochlorite ions (ClOˉ), temperature, and pH
| Accuracy | 2% or ±10 ppb TFC, whichever is greater at pH <7.5 |
| Accuracy 2 | 2% or ±10 ppb HOCl, whichever is greater at pH <8 |
| Calibration Interval | 2 months |
| Calibration Method | Zero Calibration: electrically (automatic) or with de-chlorinated water Calibration: Comparison of lab method with process sample pH Calibration: Single- or two-point using standards or comparison of lab method with process sample |
| Description | 9184sc Total Free Chlorine Amperometric Sensor |
| Enclosure Rating | IP66 / NEMA 4X |
| Flow | 14 L/h (200 – 250 mL/min) auto-regulated by flow through cell |
| Interferences | No interference from chloramines. Chlorine dioxide and ozone interfere with measurement. |
| Lower Limit of Detection (LOD) | 0.005 mg/L HOCl |
| Maintenance Interval | Measurement cell: 6 months for membrane and electrolyte for typical application pH: 1 – 1.5 years for typical application |
| Measuring Interval | Continuous |
| Measuring Principle | Amperometric/Membrane (electrode, membrane, electrolyte) |
| Mounting | Flat, vertical surface (panel, stand, etc.) |
| Operating Temperature Range | 0 – 45 °C |
| Parameter | Total Free Chlorine: Sum of hypochlorous acid, hypochlorite and free chlorine |
| pH Range | 4 – 8 pH, acidification unit available for pH >8 |
| Pressure Range | 0.1 – 2 bar in flow cell |
| Range | 0 – 20 mg/L HOCl |
| Relative Humidity | 0 – 90 % non-condensing |
| Response Time | 90 % in less than 90 seconds |
| Sample Input | 0.25 ” outer diameter |
| Sample Output | 0.5 ” inner diameter |
| Sample Temperature | 2 – 45 °C |
| Standard deviation | 0.7 |
| Storage Conditions | -20 – 60 °C |
| Temperature Compensation | Automatic over sample temperature range |
| Temperature Compensation | Automatic over sample temperature range |
| Warranty | 12 months |
| Weight | 6.5 kg |
| What’s included? | 9184sc Chlorine Sensor Panel with pH Sensor, Membranes and Electrolyte and Manual. |