A single analyzer for Carbon contamination and Nitrogen/Phosphorous nutrient levels in water
Contamination levels in water directly affect treatment and reuse decisions. By monitoring contamination and nutrient levels, water managers can make efficient, cost-effective choices for protecting and reusing valuable water supplies.
The BioTector Analyzer offers configurable solutions to meet diverse plant needs:
1. TIC & TOC (NPOC) System
Measures Total Inorganic Carbon (TIC) and Total Organic Carbon (TOC) in a liquid sample. This setup is ideal for samples with low concentrations of volatile organic compounds, delivering reliable Non-Purgeable Organic Carbon (NPOC) data.
2. TC System
Measures Total Carbon (TC), capturing the sum of TIC, NPOC, and volatile organic carbon (VOC). This comprehensive measurement helps provide a clear overview of all carbon present.
3. VOC System
Delivers measurements for TIC, TOC, TC, and VOC. Here, TOC is calculated as TC minus TIC, representing the sum of NPOC and VOC for a full picture of organic content in your sample.
With accurate and complete online monitoring of TOC, TN, and TP, you can:
• Improve treatment and reuse strategies through reliable water quality data.
• Reduce costs on nutrient dosing and protect sensitive treatment processes with precise C:N:P ratios.
• Save time on manual sampling—BioTector’s robust, industrial design ensures dependable operation, enabling you to focus on plant management.
• Detect product losses early to minimize waste and reduce the risk of noncompliance caused by accidental contamination releases.
For more details, to discuss specific application requirements, or to request additional BioTector versions, please contact us. Our team is ready to help identify the right solution for your plant.
TOC is commonly used as a fast, reliable water analysis metric for water quality. TOC levels are often correlated to lagging water quality metrics such as Chemical Oxygen Demand (COD) and Biochemical Oxygen Demand (BOD).
| Ambient Temperature | 5 – 40 °C |
| Communication | Modbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol) |
| Cycle Time | From 10 minutes, depending on range and application |
| Data Storage | Previous 9999 analysis data on screen in the microcontroller memory and storage of data archive for the lifetime of the analyzer in the SD/MMC card. Previous 99 fault data on screen in the microcontroller memory and storage of fault data archive for the lifetime of the analyzer in the SD/MMC card. |
| Display | High contrast 40 character x 16 line backlit LCD with LED backlight |
| Drain Pressure | Typically ambient (for applications with high drain pressure, optional systems are available) |
| EExp / Hazardous Location | Certification options are available to European Standards (ATEX Zone 2 – maximum T3 for TP analyzer) and North American Standards (Class I Division 2) |
| Expected Measuring Range | 0 – 25 mg/L |
| Humidity | 5 – 85 % (non-condensing) |
| Languages | English |
| Lower Limit of Detection (LOD) | TOC: 0.6 mg/L C with Automatic Range Selection TN: 0.4 mg/L N with Automatic Range Selection TP: 0.4 mg/L P with Automatic Range Selection |
| Measurement Method | TOC: NDIR measurement of CO₂ after oxidation TN: direct photometric analysis of Nitrate after oxidation TP: colorimetric analysis of Phosphate with standard Vanadomolybdophosphoricacid method after oxidation |
| Multi-Stream | Valves for up to 3 streams with up to six 4-20 mA signals The number of available outputs depends on the manual stream configuration. |
| Number of Channels | 1 Channel (One 4-20mA output is indluded as standard) |
| Outputs | One programmable 4-20 mA analogue output signal (typically for TOC) Maximum impedance: 500 ohms For systems requiring more than six 4-20 mA standard outputs, 4-20 mA Output Multiplex option is implemented to provide 4-20 mA data for up to 35 output signals |
| Oxidation Method | Patented Two-Stage Advanced Oxidation Process (TSAO) using Hydroxyl Radicals |
| Parameter | Direct measurement of TOC, TIC, TC, TN, TP; COD, BOD via correlation; VOC via calculation |
| Particle Size | Up to 2 mm, soft particulates |
| Permissible Chloride Range | Up to 30% |
| Power Requirements (Hz) | 50/60 Hz |
| Power Requirements (Voltage) | 115 VAC |
| Power Supply | TIC/TOC & TN/TP – 115V |
| Range | 0 – 25 mg/L C/N/P |
| Range Selection | Automatic or Manual Range Selection |
| Repeatability | TOC: ±3% of reading or ±0.3 mg/L C, whichever is greater TN: ±3% of reading or ±0.2 mg/L N, whichever is greater TP: ±3% of reading or ±0.2 mg/L P, whichever is greater |
| Sample Inlet Temperature | 2 – 60 °C |
| Sample Volume | Up to 8.0 mL |
| Service Interval | 6 months service intervals |
| User Interface | Microcontroller with membrane keyboard |
| Warranty | 12 months |
| Warranty | 1 |
| Weight | 90 – 120 kg Enclosure weight may change depending on system optional features. |
| What’s included? | B7000 TOC/TN/TP Analyser, Tubing, Fuses, Ferrules, Drain, Acid, Base & TN Dip Tubes, CO₂ Filter & B7000 TOC/TN/TP User Manual |