Technologies are advancing rapidly, providing new levels of convenience, accuracy, and efficiency. Which is exactly why the SC4500 Controller from HachⓇ is designed to integrate easily into your current system while allowing you to upgrade as your capabilities advance, without having to replace inventory. With a wide range of analog and digital connectivity options and the availability of intelligent instrument and data management features, the SC4500 unlocks the future, today.
| Altitude | 3000 m (9842 ft) maximum |
| Analog Output Functional Mode | Linear, PID |
| Communication (optional) | Analog: Five 0 – 20 mA or 4 – 20 mA analog outputs on each analog output module Up to two analog Input modules (0 – 20 mA or 4 – 20 mA). Each input module replaces a digital sensor input. Digital: |
| Compatible Instruments | Compatible Sensors and Analysers / Software Version (Release Year) Amtax sc / V2.30 (2018) or higher A-ISE sc / V1.02 or higher AN-ISE sc / V1.08 (2013) or higher N-ISE sc / V1.02 or higher NH6000sc Ammonium Analyser Nitratax clear sc, Nitratax eco sc, Nitratax plus sc / V3.13 (2013) or higher NT3100sc/NT3200sc Phosphax sc / V2.30 (2018) or higher Phosphax sc LR/MR/HR / V1.01 (2018) or higher TSS sc / V41.73 (2013) or higher Solitax sc / V2.20 (2013) or higher TU5300sc, TU5400sc / V1.34 (2017) or higher SS7 sc (in Bypass) / V1.01 (2006) or higher Ultraturb sc / V3.06 (2017) or higher 1720E / V2.10 (2006) or higher Sonatax sc / V1.15 (2016) or higher CL17sc / V2.7 (2019) or higher CL10sc / V1.14 (2013) or higher 9184sc, 9185sc, 9187sc* / V2.03 (2013) or higher Uvas plus sc / V3.01 (2017) or higher LDO 2 sc* / V1.22 (2013) or higher 3798sc* / V2.03 (2013) or higher 3700sc + Inductive Conductive Digital 6120800 / V3.00 (2017) or higher 3422sc + Contacting Conductive Digital 6120700 / V3.00 or higher 3700 analog + Conductivity Module LXZ525.99.D0004 3400 analog + Conductivity Module LXZ525.99.D0004 pHD sc*, pHD-S sc / V3.10 (2016) or higher 1200-S sc* / V2.04 (2013) or higher pHD analog + Digital Gateway 6120500 / V3.00 (2017) or higher pHD analog + pH/ORP Module LXZ525.99.D0003 RC and PC analog sensor + Digital Gateway for conventional analog pH and ORP sensors 6120600 / V3.00 (2017) or higher RC and PC analog + pH/ORP Module LXZ525.99.D0003 8362sc* / V3.00 (2017) or higher Polymetron pH/ORP analog + Ultrapure pH/ORP Module LXZ525.99.D0007 Polymetron Conductivity analog + Ultrapure Conductivity Module LXZ525.99.D0006 GS1440 and GS2440EX Sensors H₂S FP360 sc / V1 or higher *Hardware Version1 of instrument is not supported |
| Compatible Network Technologies | GSM 3G/4G (e.g. AT&T, T-Mobile, Rogers, Vodafone etc.) CDMA (e.g. Verizon) |
| Compliance Certifications | CE. ETL certified to UL and CSA safety standards (with all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C م (Morocco) |
| Conduit Openings | ½” NPT conduit |
| Description | Microprocessor-controlled and menu-driven controller that operates the sensor |
| Dimensions | ½ DIN – 144 x 144 x 192 mm (5.7 x 5.7 x 7.6 in.) |
| Display | 3.5-inch TFT colour display with capacitive touchpad |
| Enclosure Rating | UL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X Metal enclosure with a corrosion-resistant finish |
| Indoor/Outdoor | Outdoor installation in direct sunlight or UV radiation requires UV protection screen and/or sunroof |
| Installation Category | Category II |
| Material Enclosures | Polycarbonate, aluminum (powder coated), stainless steel |
| Measurements | Two device digital SC connectors |
| Mounting | Wall, Pole, or Panel Mounting |
| Network Connectivity | Optional: LAN: Two Ethernet connectors (10/100 Mbps) Cellular: External 4G Wi-Fi |
| Operating Temperature Range | -20 – 60 °C (-4 – 140 °F) (8 W (AC)/9 W (DC) sensor load) -20 – 45 °C (-4 – 113 °F) (28 W (AC)/20 W (DC) sensor load) Linear derating between 45 and 60 °C (-1.33 W/°C) |
| Outputs | Ethernet IP |
| Pollution Degree | 4 |
| Power Requirements (Voltage) | 18 – 28 VDC; 2.5 A |
| Power Supply | Without plug 24 VDC |
| Protection Class | I, connected to protective earth |
| Relays | Two relays (SPDT); Wire gauge: 0.75 – 1.5 mm² (18 – 16 AWG) AC controller DC controller Optional Maximum switching voltage: 30 VDC |
| Sensor Input #1 | Digital |
| Sensor Input #2 | mA Input |
| Software available | Non-Claros |
| Storage Conditions | -20 – 70 °C (-4 – 158 °F), 0 – 95% relative humidity, non-condensing |
| USB Port | Used for data download and software upload. The controller records approximately 20,000 data points for each connected sensor. |
| Warranty | 12 months |
| Weight | 1.7 kg (3.7 lb) (controller only, w/o modules) |
| What’s included? | Controller, Non-Claros + Embedded Prognosys, without power cord; includes mounting hardware |