The first maintenance-free optical oxygen sensor for nuclear power plants.
The Hach Orbisphere K1200 sensor is a high accuracy Luminescent Dissolved Oxygen (LDO) sensor that has been optimized for measurements in water processes in the power and steam industry.
Designed for mechanical robustness to extend service life and optimize total cost of ownership, the sensor, combined with the Orbisphere Controllers, offers a new way to monitor dissolved oxygen in power plants. Based on good engineering practice the Orbisphere K1200 LDO sensor is suitable for nuclear plant application.
Maintenance and operating costs are significantly reduced due to the absence of membrane and electrolyte and sensor accuracy is unaffected by process changes such as flow changes.
| Accuracy | ± 0.8 ppb or 2 % whichever is greater |
| Ambient Temperature | -5 – 50 °C (23 – 122°F) |
| Analog Outputs | 3 Smart 0/4 to 20 mA (500 Ohms) programmable as linear or tri-linear, configurable to send diagnostics or alarm informations. |
| Application | Nuclear |
| Calibration | Single point zero calibration with standard 99.999% nitrogen (quality 5.0) or equivalent oxygen free gas |
| Certifications | 2004/108/EC – EN 61326-1 |
| Communication Capabilities | 3 x 0/4-20 mA; RS485; Ethernet |
| Display Resolution | 0.1 ppb |
| Lower Limit of Detection (LOD) | 0.6 ppb |
| Mounting | 28 mm Orbisphere fitting |
| Parameter | Oxygen |
| Power Requirements (Voltage) | 100 – 230 VAC |
| Range | 0 – 2000 ppb dissolved O2 (DO) |
| Relays | Measurement board: 3 measurement alarm relays (1A-30 VAC or 0.5A-50 VDC), configurable to send diagnostics information. Main board: 1 system alarm relay (1A-30 VAC or 0.5A-50 VDC). |
| Repeatability | ± 0.4 ppb or 1 % whichever is greater |
| Reproducibility | ± 0.8 ppb or 2 % whichever is greater |
| Response Time | (90%) <10 s (gas phase); <30 s (liquid phase) |
| Sample Flow Rate | 50 – 300 mL/min |
| Sample Pressure | 1 – 20 bar absolute (14.5 – 290 psi) |
| Sample Temperature | -5 – 50°C (23 – 122°F) |
| Sensors | K1200 LDO |
| Storage Conditions | -5 °C – 100 °C |
| Temperature Range | Accurate from -5 – 50 °C (23 – 122 °F) Resistant – temperature from -5 – 100 °C (23 – 212 °F) |
| Warranty | 12 months |
| Weight Sensor | 0.6 |
| What’s included? | K1200 Sensor only. Controller 510, sensor cable, flow-chamber or process connection must be ordered separately. |