An integral part of the most complete water analytics system for the Power industry. Hach provides a broad range of product options designed to work together into flexible solutions to meet your unique needs. Hach’s comprehensive approach saves you time on design, installation, training, maintenance, and operation.
| Analog Outputs | Two (Five with optional expansion module) 0/4 – 20 mA isolated current outputs, max. 550 Ω, Accuracy: ±0.1% of FS (20 mA) at 25 °C, ±0.5% of FS over -20 °C to 60 °C range |
| Cable Length | 10 m (33 ft) |
| Calibration Method | Zero: Electrically or with oxygen free water, Slope: in air or against a laboratory measurement |
| Communication | MODBUS RS232/RS485, PROFIBUS DPV1, HART optional |
| Communication Capabilities | Profibus DP (includes sculpted barrier) |
| Conduit Openings | 1/2″ NPT Conduit |
| Connection Drain Line | 8 mm tubing |
| Connections | 1/4″ NPT thread (6mm or 1/4″ tubing advised) |
| Display | Graphic dot matrix LCD with LED backlighting, transreflective |
| Display Resolution | 240 x 160 pixels |
| Display Size | 48 x 68 mm |
| Electrical Certifications | EMC CE compliant for conducted and radiated emissions: – CISPR 11 (Class A limits) – EMC Immunity EN 61326-1 (Industrial limits) Safety CAN/CSA C22.2 No. 61010-1 |
| Enclosure Rating | IP66 / NEMA 4X |
| Lower Limit of Detection (LOD) | < 1 ppb |
| Maintenance Interval | Membrane Lifetime: 6 months depending on sample |
| Manual Languages | Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish |
| Material | Electrodes: Gold cathode, Silver anode; Flow-through cell: Stainless Steel 316L; Membrane: PFA, Sensor Body: Noryl |
| Material Enclosures | Polycarbonate Aluminium (powder coated) Stainless Steel |
| Measurements | 0 to 2000 ppb (0-2 ppm) |
| Operating Temperature Range | -20 – 60 °C at 0 – 95% RH (non-condensing) |
| Power Requirements (Hz) | 50/60 Hz |
| Power Requirements (Voltage) | 100 – 240 VAC |
| Range | 0 – 2000 ppb (0-2 ppm) |
| Range | 0 – 2000 ppb |
| Relay Functions | Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning |
| Relays | Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A |
| Repeatability | ± 0.5 ppb or ±5%, whichever is greater |
| Reproducibility | ± 0.5 ppb or ± 2% whichever is greater |
| Response Time | For step change 1-40 ppb: <30s |
| Sample Flow Rate | 66 – 166 |
| Security Levels | 2 |
| Storage Conditions | -20 °C – 70 °C |
| Temperature Compensation | Automatic in the range of 0 – 45 °C (32 – 113 °F) |
| Temperature Compensation | Automatic in the range of 0 – 45 °C (32 – 113 °F) |
| Units | mg/L, ppm, µg/L, ppb, mbar, hPa, inch Hg, mmHg |
| Warranty | 12 months |
| Weight | 3.2 kg |
| What’s included? | Controller, Mounting Hardware, Oxygen sensor, Temperature Sensor, Cable, Flow-Chamber, Basic User Manual |