Simple to install. Safe to operate.
Hach’s Sample Conditioning Panel (SCP) for a single analyzer allows any analyzer to be easily connected to a hot and pressurized sample. It helps to protect and respect all safety and sample requirements.
| Application | Steam sample only in low pressure boilers for Power & Industry |
| Cooling Water | Filtered, softened/decarbonized water quality required |
| Permissible Chloride Range | <250 mg/L for sample temperature 77 – 356 °F (25 – 180 °C) <100 mg/L for sample temperature 356 – 554 °F (180 – 290 °C) <25 mg/L for sample temperature 554 – 1022 °F (290 – 550 °C) |
| pH Range | 7 – 12 pH |
| Pressure Range | Cooling water: <150 psig (10.3 bar) |
| Sample Flow Rate | 200 mL/min for conductivity 150 mL/min for SiO₂ and PO₄ 100 mL/min for pH, O₂, Na Plus 350 mL/min for grab sample Maximum total sample flow for the panel is 1600 mL/min for hot water or 1000 mL/min for steam. |
| Sample Pressure | <300 psig (20.7 bar) |
| Sample Temperature | <600 °F (315 °C) |
| Temperature | Cooling water: <110 °F (43 °C) |
| Turbidity | <50 NTU |
| What’s included? | Sample Conditioning Panel for online analysis, consisting of the following components: Sentry Back Pressure Relief Valve with a Set-point of 20 psig (1.4 bar), Grab Sample, 304 SS Sample Sink; all components are factory mounted on a 12 in. (W) x 36 in. (H), T304SS backplate. All sample tubing throughout panel is SA249, T316 SS. Cooling Water Valve Materials are Stainless Steel. Sentry Thermal Shut off Valve with a Set-point of 120 °F (49 °C). Low Pressure (Cooled Sample) Blow-down Valve, Total Flow Indicator; 0-1600 cc/min, Temperature Gauge; 0-250 °F (-20-120 °C) Pressure Gauge; 0-60 psig (0-4 bar). Globe Valve on the Cooling Water Outlet. 3-Way Cooling Water Inlet Valve with drain. 3/8 in. Sample Inlet Isolation Valve. Primary Sample Cooler, Sentry Model FLR6222. Sample Flow Control Valve. |