Outstanding results in high temperatures: Surface Scatter 7 sc
Able to measure very high and low turbidity levels accurately, this turbidimeter is uniquely designed so that the light source and the photocells never come in contact with the sample.
The sample is introduced into the center sample tube of an inclined turbidimeter body at a flow rate of 1 – 2 liters per minute. As the fluid spills over the top of the turbidimeter body, a stable, flat surface of liquid forms and becomes the measuring surface.
The light source emits a high-intensity beam of light, which strikes the surface of the fluid diagonally. The detector registers the light that is scattered at 90° to the incident beam. This scattered light is proportional to the content of suspended particles in the sample.
| Accuracy | ±5% of reading or ±0.1 NTU (whichever is greater) from 0 – 2000 NTU |
| Accuracy 2 | ±10% of reading from 2000 – 9999 NTU |
| Alarm | Relays can be programmed through sc controller for warning, alarm, feeder, event, pulse width modulation, and frequency control. |
| Ambient Temperature | 0 – 50 °C |
| Channels | Instrument Only |
| Controller | No |
| Controller Compatibility | SC4500, SC1000, SC200 |
| Enclosure Construction | Plastic instrument enclosure |
| Enclosure Rating | NEMA 12 |
| Installation Category | Category II |
| Mounting | Wall mount |
| Number of Channels | None |
| Power Requirements (Voltage) | 100/230 VAC |
| Range | 0.01 – 9999.9 NTU |
| Relative Humidity | 5 – 95% non-condensing |
| Repeatability | 1.0% or ±0.04 NTU, whichever is greater |
| Resolution | <1000 NTU: 0.01 NTU 1000 – 9999.9 NTU: 0.1 NTU |
| Response Time | Initial response in 45 seconds |
| Sample Flow Rate | Sample flow required 1.0 – 2.0 L/min |
| Sample Input | 0.75 ” NPT female thread |
| Sample Output | 1 ” NPT female thread |
| Sample Temperature | 0 – 50 °C |
| Storage Conditions | -20 – 80 °C |
| Warranty | 24 months |
| Weight | 15.8 kg (34.8 lb) |
| What’s included? | SS7®sc, Wall Mount Kit, Cylinder Brush, Light Source Alignment Template, 0.75 x 0.75 inch NPT Nipple, Calibration Cup, 0.75 inch NPT Ball Valve, SS7sc User Manual, and 500 mL 4000 NTU Formazin Standard. |

Tiếng Việt