The turbidity bypass sensor that brings clarity
Turbidities in the range of FNU (NTU) 0.0001-1,000 are measured and displayed alphanumerically using a controller. This high resolution facilitates precision differentiation, even for ultra clear liquids. A pulsed, long-life IR radiation source (LED) ensures that costs are minimized. Designed in accordance with DIN EN ISO 7027 and permanently calibrated before leaving the factory. All key data is set to practical standard values. The instruments are immediately ready for use after connection to the power supply and the supply of sample water. The measuring range and all data outputs are modified to suit requirements using menus on the controller. All optical and electronic assemblies are installed in housings that are physically very strong and proof against water jets.
| Accuracy | ± 0.008 FNU ±1% of reading whatever is greater (0 – 10 FNU) |
| Automatic Wiper | Yes |
| Cable Length | 30 m |
| Calibration Method | Stablcal or dry standard CVM module |
| Dimensions (H x W x D) | 240 mm x 250 mm x 110 mm |
| Display | Hach SC 200 Digital Controller or Hach
SC 1000 Universal Controller |
| Inlet | 13 mm ID Tubing or fixed connection (PVC system component pipework) |
| Material | Housing: ASA; Wiper Blade: Silicon; Wiper Shaft: Stainless steel 1.4571 |
| Measurement Method | 90° infrared pulse scattered light technique in accordance with DIN EN ISO 7027 |
| Model | Ultraturb seawater sc |
| Operating Temperature Range | -10 – 40 °C (Display Unit) |
| Outlet | 13 mm ID Tubing or fixed connection (PVC system component pipework) |
| Pressure Range | 6 bar at 20 °C (87 psi at 68 °F) |
| Range | 0.0001 – 1000 FNU freely programmable (0.0001 – 250 EBC = 2500 ppm SiO2) |
| Repeatability | ± 0.003 FNU or ± 0.5% of reading (whichever is greater) from 0 – 2 FNU |
| Response Time | 1 – 60 s (programmable) |
| Sample Flow Rate | 0.2 to 1 L/min |
| Sample Temperature | Max. 50°C (122°F) |
| Sensor cleaning | Wiper, automatic |
| Verification | Stablcal or dry standard CVM module |
| Weight | Approx. 1.5 kg |
| What’s included? | Hach Ultraturb sc, Factory Test Certificate, 4 wipers, Accessory set LZP816, User Manual. |